HI-Q Digital Air Flow Calibrator D-AFC

Brand:HI-Q
model:D-AFC

If you want to learn about the product, please contact us.
Please provide the brand, model, and quantity for inquiry.

Category:

Description

Product Highlights

  • Flow Rate – Instantaneous Display of actual & standard flow rate (ACFM, ALPM, ACMH, SCFM, SLPM, SCMH)
  • Flow Totalizer – Instantaneous Display of total flow in actual & standard units (ACF, AL, ACM, SCF, SL, SCM, SCMH)
  • Temperature Sensor- Measurement & display of temperature of inline air flow (degrees F & C)
  • Barometric Pressure Sensor – Measurement & display of barometric pressure (in Hg & mm Hg)
  • Auto Calibration of HVP-4000 & 5000 Series Air Sampler
  • Display – Illuminated graphic LCD screen
  • Data Logger – Internal memory for data logging. Data downloadable to PC/Laptop computer.
  • Data Output – Optional analog 4-20 mA or
    0- 10 VDC scaled output
  • Printers – Optional Data / Barcode Printers

Specifications

Parameter Display Range Resolution
Barometric pressure: 0-32 in Hg;
0-812 mm Hg
0.01 in Hg
0.01 mm Hg
Temperature: 30-900 deg F
-1-480 deg C
0.01 deg F
0.1 deg C
Air Flow: See “Calibration Range” under “Ordering Information”
Measurement Accuracy: Flow Rate: ± 2% of full scale
Temperature: ± 2 deg F
Barometic Pressure: ± 0.2 in Hg @ 25 deg C
Flow Display: cfm, lpm or cmh (can be programmed for any other units)
Power Requirements: 110 – 240 VAC
Overall Dimensions: 12” W X 6” D X 11” H
Weight: 8.5 lbs
Storage Temperature: -4ºF to 158ºF (-20°C to + 70°C)
Operating Temperature: 14ºF to 140ºF (-10°C to + 60°C)
Calibration: Factory calibration is recommended once per year

Through the implementation of a baro-metric pressure sensor, thermocouple, precision machined venturi tube and differential pressure sensor, HI-Q’s D-AFC- Series microprocessor based air flow calibrators are used to measure, record, & display instantaneous flow rate (i.e.: CFM, LPM, or CMH), temperature (degrees C or F), and barometric pressure (“Hg or mm Hg). Through onboard calculations the user may also select to display the flow rate or total volume in “Standard Units” (ie: SCFM, SLPM, SCMH or SCF, SL, SCM). The total volume of air sampled is displayed (resettable) in standard and actual units. The processor also has an internal memory that can be programmed for logging and time stamping the instantaneous flow rate, barometric pressure, and temperature. The data can be downloaded to an excel spread-sheet using a PC or a laptop computer. Networking and communications options include two (selectable) RS-232/RS-485 ports, a 4-20 mA and/or 0-10 VDC analog output (proportional to flow rate), the ability to send & receive SMS messages to/from any CDMA/GSM cellular phone (for possibly alerting/reporting any predefined event via text message), remote or local data acquisition and the ability to print in textual and bar code format.

Buyer Notice

1. About the product

Due to market fluctuations, exchange rate changes, and other factors, product prices and product attributes are for reference only; The specific product requirements are subject to customer requirements, and the prices are subject to communication and quotation. All product prices on this website are not considered transaction prices and are for reference only!

2. Regarding trademark copyright

All company names, trademarks, logos, product images, product descriptions, product brand names, etc. related to this website, and all intellectual property rights such as trademark copyright belong to the corresponding brand; This website is only for the convenience of product identification.
Please carefully read the buyer’s instructions to avoid unnecessary disputes.

Reviews

There are no reviews yet.

Be the first to review “HI-Q Digital Air Flow Calibrator D-AFC”