The Oilless Rotary Vane, Air Sampling Pumps incorporated in HI-Q’s continuous duty air sampling systems have “Lubricated-for-Life” ball bearings which leaves the internal workings virtually maintenance free. The only servicing needed is the occasional replacement of vanes, “O” Rings,
filters, and gaskets when they become worn. Many variables determine the life expectancy of a pump, such as ambient temperature, operating cycle/speed, duty level, condition of air handled and unit maintenance. HI-Q recommends inspecting the pump system every 2,000 hours. Inspection is done within a few minutes by removing a few bolts and the pump end casting which exposes the rotor and vanes.
ご注文について
| System Model Number | XXX-0211CRNV | XXX-0523CV | XXX-1023CV |
|---|---|---|---|
| Replacement Motor/Pump | RV11-0211 | RV23-0523 | RV23-1023 |
| Service Kits (Includes 4 Vanes) | K218 | K478 | K479 |
| Single Replacement Vanes (4 Required) | AA348A | AH850A | AK513 |
| Filter/Muffler End Cap Assembly | AA617G | AK526 | AK526 |
| 8 Foot 14/3 Grounded Power Cord | AA816 | AA816 | AA816 |
– 0523 & 1023 Service kits include: 4 Carbon Vanes, 2 Felt Filters, 1 Motor Gasket and 2 Felt Filter End Cap Assembly “O”-Rings.
– 0211 Service kits include: 4 Carbon Vanes, 1 Body Spacer, 2 Muffler Jar Cover Gaskets, & 4 Felt Filters.