描述
The TPI 709R Flue Gas Analyzer is a state of the art, easy to use analyzer designed not only to display and calculate the required readings from a flue but also to cover most of the other measurements associated with combustion.The instrument is ruggedly constructed and comes with a 3 Year unitand 2 Year sensor Guarantee.
The TPI 709R combustion analyzer uses state of the art electrochemical sensors. This sensor technology provides the longest lasting, most accurate and reliable means for performing combustion tests. The sensors in your analyzer will need to be replaced periodically and calibration is recommended once every year. The 709R will remind you when calibration is due by flashing “CAL DUE Err” for three seconds at start up.
Electrochemical sensors by nature are always active. Therefore the time the analyzer is off and not being used must be taken into account when determining sensor life. The sensors in your analyzer are warranted for two years. This warranty does not cover sensors damaged through misuse of the analyzer.
You should keep the batteries of your 709R fresh so power is constantly being supplied to your sensors.
The following guidelines will help prevent damage to your sensors:
- Always use the mini pump filter when testing flue gases.
- Always periodically check and replace the mini pump filter as needed.
- Always make sure the in-line filter / water trap is installed properly.
- Always periodically check and replace the in-line filter as needed.
- Always remove water or condensation from the inside of the in-line filter / water trap assembly prior to performing tests.
- Always use the optional oil filter (p/n A773) when performing tests on oil burning equipment.
- Never over saturate your sensors by performing tests on equipment with gas levels beyond the capability of you analyzer.
- Always keep the A794 water trap / filter assembly clean and replace the filter as necessary. Replacement filter part number is A794F.
The TPI 709R Flue Gas Analyzer comes complete with the following standard accessories:
- TPI 709R Instrument
- Rubber Boot (A765)
- Battery Charger (A766) – 1 each
- Soft Carrying Case (A768)
- Flue Sampling Probe (A770)
- In-Line Filter assembly installed on Flue probe (A794)
- Temperature Probe (GK11M)
- Mini Pump Protection Filter Assembly and spare filters (A763)
- Exhaust Spigot (removable) (A764)
- Instruction Manual
The TPI 709R Flue Gas Analyzer has the following options available:
- Infrared printer (A740)
- Spare In-Line Filter (A794F is a package of 5 filters)
| Model 709R Features | ||
|---|---|---|
| Measure CO, O2 & Flue Temperature | ||
| Calculate CO2, CO/CO2 Ratio, Efficiency & Excess Air | ||
| Dual Input Differential Manometer – Great for performing Standing Pressure or Let By/Tightness Tests | ||
| Dual Input Differential Thermometer – Ideal for measuring & recording Flow & Return Temperatures | ||
| Ambient CO Analysis | ||
| 6 Year Warranty* (* subject to annual service being carried out each year by TPI or one of its approved local service centres) | ||
| Diagnostic software & technical helpline combine to save time and money on reduced downtime as many of the reported faults can be rectified over the phone keeping you and your instrument working | ||
| Unique external water trap filtration system as standard on all kit options preventing loss of downtime or additional costs due to potentially damaged sensors caused by water ingress | ||
| 10 Fuel options including Wood, Light & Heavy Oil, LPG and Natural Gas | ||
| Will not shut off if CO ppm is above 15ppm for increased safety for both the user and the instrument | ||
| Unique in-line pump protection filter fitted as standard across all kits preventing loss of downtime or additional costs due to a potentially damaged pump caused by dust/dirt ingress | ||
| Protective rubber boot as standard to protect the FGA | ||
| Re-Chargeable battery housed in a separate compartment ensuring no heat transfer when using on the mains to affect gas readings | ||
| All kits come supplied in a soft carry case to protect not only the FGA but all of your accessories | ||
| Capability of sending Live or Stored readings to an optional IR Printer | ||
规格
| Model 709R Specifications | ||
|---|---|---|
| 仪器 | ||
| 工作温度范围 | 14°F 至 +122°F (-10°C 至 +50°C) | |
| Battery / Batery Life | Rechargeable Ni-MH / > 6 Hours | |
| Charger Input Voltage | 115V or 230V : 50/60 Hz AC | |
| 燃料 | Natural Gas, LPG, Light Oil, Heavy Oil, Bituminous Coal, Anthracite Coal, Coke, Butane, Wood, Bagasse |
|
| 压力范围 | mbar, kPa & inH2O | |
| 显示屏 | 背光图形 LCD | |
| 数据存储 | 50 sets of readings, multiple pages | |
| 时间和日期 | 24 小时实时时钟 | |
| 尺寸 | 7.8” (200mm) x 3.5” (90mm) x 2.4” (60mm) | |
| 重量 | 500g | |
| Casing | Rubber Boot as Standard | |
| Switch Off | Failsafe | |
| Exhaust | Safety Spigot | |
| 符合 | EN50379 | |
| 烟道温度探头 | ||
|---|---|---|
| 建筑 | 手枪式握把与不锈钢轴 | |
| 软管长度 | 8.2 英尺(2500 毫米) | |
| 插入长度 | 7.9 英寸(200 毫米) | |
| ‘K’ Type Thermocouple Accuracy | 全量程 +/- 0.3%, +/- 2°F (1°C) | |
| 最高温度 | 1472°F (800°C) | |
| 气体 | 范围 | 决议 | 准确性 | |
| 氧气 | 0-25% | 0.1% | +/- 0.3% | |
| Carbon Monoxide | 0-10,000 ppm | 1 ppm | (<20ppm) +/- 3 ppm (>=100ppm) +/- 5% |
|
| Carbon Dioxide (calculated | 0-25% | 0.1% | 计算 | |
| CO/CO2 比率 | 0-0.999 | 0.001 | +/- 0.3% | |
| 燃烧效率 | 0-100% | 0.1% | ||
| 可选范围 | mbar, kPa and inH2O | |
| 范围 | – 150 mbar to + 150 mbar | |
| 决议 | 0.01 mbar & 0.1 mbar | |